Agilent Technologies 6850 Series II User Information page 157

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Table 30. PTV Inlet liners
Type
Open baffled liner
Liner packed with
silanized glass wool
Unpacked liner, to be
packed by the user
Type
Single baffle liner
Single baffle liner
Multi baffle liner
Fritted glass liner
*Silanized glass wool 10 gm (pesticide grade) part no. 5181-3317
Replacing liners
1.
2.
Released: March 2004
Injection capacity
Lowest capacity
Higher capacity
Depends on the packing
Injection
capacity
180 µL
200 µL
150 µL
150 µL
Remove the head from the inlet. See
"Removing the septum
Grasp the liner by the GRAPHPACK ferrule. Remove the liner and ferrule.
6850 Series II GC User Information
Inertness
Most inert
Less inert
Glass wool
Glass type
packing*
Borosilicate
Yes
deactivated
Borosilicate
No
deactivated
Borosilicate
No
deactivated
Borosilicate
No
deactivated
"Removing the septumless head"
head".
Routine Maintenance: Inlets
Maintaining a PTV
Quantity
Part no.
10
5182-9751
10
5182-9752
10
5182-9753
Typical
application
Part no.
Large volume
5183-2038
injection, not for
extremely active
compounds
General purpose
5183-2036
Active com-
5183-2037
pounds, drugs,
pesticides
Large volume
5183-2041
injection, all but
the most active
compounds
page 157 of
or
256

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents