Safely route vent tubing

The FPD provides a plastic vent tube that exits the back of the detector cover.
During normal operation, this detector produces water vapor. Vent the tubing
appropriately.
The µECD provides a vent tube that exits the back of the lid. Connect this tube
to a fume hood or other location as directed by your radiation safety officer or
environmental health and safety organization. See
for more information.

Prepare the cryogenic oven cooling system

Warning
Pressurized liquid CO
personnel from high pressures and low temperatures. CO
concentrations is toxic to humans; take precautions to prevent hazardous
concentrations. Consult your local supplier for recommended safety
precautions and delivery system design.
Warning
Do not use copper tubing or thin-wall stainless steel tubing with liquid CO
Both harden at stress points and may explode.
Caution
Liquid CO
because the expanding liquid may form solid CO
dry ice builds up in the oven, it can seriously damage the GC.
If you ordered CO
oven and the filter on the cryogenic chassis are shipped uninstalled because of
the possibility of damage if shipped installed.
Released: March 2004
is a hazardous material. Take precautions to protect
2
should not be used as a coolant for temperatures below –40°C
2
cryogenic oven cooling with the GC, the cooling coil for the
2
6850 Series II GC User Information
Safely route vent tubing
"µECD warnings" on page 7
in high
2
—dry ice—in the GC oven. If
2
Installation
.
2