3.
4.
5.
6.
7.
Released: March 2004
Hold the adapter in this position and tighten the nut finger tight. Tighten
1/4 turn more with a wrench.
Place the ferrule and column nut on the column end. Trim off a short piece
to remove any ferrule fragments inside the column.
Gently insert the column into the detector until it bottoms; do not attempt
to force it.
Tighten the column nut finger tight. Withdraw the column about 1 mm.
Use a wrench to tighten the nut an additional 1/4 turn.
6850 Series II GC User Information
Columns and Traps
Installing capillary columns
page 30 of
256