Vial stripper arm
6.
7.
8.
Installing the turret
This is the reverse of the removal process just described.
1.
2.
3.
Released: March 2004
To remove the turret, lift the turret straight up to free it from the motor
hub.
Tilt the near edge of the turret up and pull it out slowly up and toward you
until it clears the roller on the vial stripper arm. If you move it too far
toward you it may get caught on the motor hub.
Rotate clockwise to clear the stripper, then simultaneously raise and
rotate counterclockwise to clear the injector body. Remove the turret.
Note that the bottom of the turret can easily get caught on the motor hub.
If you have the vials removed, you can look through the empty vial
positions and see if the inside edge of the turret is caught on the hub.
Orient the turret so that the cutout in the turret faces the injector.
Place the turret over the hub so that the metal lip on the turret bottom is
past the hub, then tilt the far end down. Slip the edge of the turret between
the hub and the edge of the injector body. Gently push it inwards and
down till it rests on the hub.
Rotate the turret on the hub until the arrow marked on the turret aligns
with the circular depression on the hub. The turret will now seat flat
against the hub.
6850 Series II GC User Information
Automatic Liquid Sampler and Injector