It is unlikely, even in this very unusual situation, that radioactive material will
escape the cell. However, permanent damage to the
is possible, and therefore, the cell must be returned for exchange.
Warning
Do not use solvents to clean the µECD.
You may not open the µECD cell unless authorized to do so by your local
nuclear regulatory agency. Do not disturb the four socket-head bolts. These
hold the cell halves together. Removing or disturbing them is a violation of the
terms of the General License and could create a safety hazard.
Safety precautions when handling µECDs
Always observe the following precautions:
Agilent Technologies recommends a vent line inside diameter of 6 mm
(1/4-inch) or greater. With a line of this diameter, the length is not critical.

Safety and regulatory certifications

The 6850 GC conforms to the following safety standards:
Released: March 2004
After turning off the main power and allowing the instrument to cool, cap
the detector inlet and exhaust vent openings. Wear disposable plastic
gloves and observe normal laboratory safety precautions.
Return the cell for exchange. Contact your local sales office for details.
Include a letter stating the condition of abuse.
Never eat, drink, or smoke when handling µECDs.
Always wear safety glasses when working with or near open µECDs.
Wear protective clothing such as laboratory jackets, safety glasses, and
gloves, and follow good laboratory practices. Wash hands thoroughly with
a mild non-abrasive cleaner after handling µECDs.
Cap the inlet and outlet fittings when the µECD is not in use.
Connect the µECD exhaust vent to a fume hood or vent it to the outside.
Canadian Standards Association (CSA): C22.2 No. 1010.1
CSA/Nationally Recognized Test Laboratory (NRTL): UL 61010A–1
International Electrotechnical Commission (IEC): 61010–1
EuroNorm (EN): 61010–1
6850 Series II GC User Information
Important Information
Safety and regulatory certifications
63
Ni plating within the cell