4.
5.
6.
7.
Released: March 2004
Disconnect the ignition assembly at the cable connection. Squeeze the
lock and gently pull the connector free.
Ignition wire
Unscrew the ignition wire from the detector. Do not lose the small copper
washer underneath it.
Install the new ignition wire.
Restore normal operating conditions.
6850 Series II GC User Information
Routine Maintenance: Detectors
Maintaining a flame ionization detector (FID)
page 199 of
256