Minimum
source
pressure
(psig)
Figure 23. Pressure requirements for auxiliary EPC flow restrictors
Released: March 2004
Pressure requirements for auxiliary EPC flow restrictors
(ambient conditions: 25°C, 14.7 psia)
80
70
19234-60660
60
50
40
30
20
10
0
0.1
1.0
with air, nitrogen, and helium
6850 Series II GC User Information
with air, nitrogen, or helium
(blue dot)
19231-60770
(red dot)
10.0
Gas flow in mL/min
Routine Maintenance: General
Maintaining the auxiliary EPC module
19231-60610
(brown dot)
100.0
page 117 of
1000.0