Setting Your Ip Address - Agilent Technologies 6850 Series II User Information

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Setting your IP address

Each of the three LAN cards that could be installed in your 6850 GC provide
different methods of obtaining the GC's IP address.
LAN cards and their available control modes.
Table 38. Available Control Modes
LAN card
J2552B
J4100A
Lantronics
a Refer to your Cerity Chemical or ChemStation software CD for a copy of BootP server software
b Local addressing requires LAN card firmware revision K.08.04 or greater
To determine which LAN card is installed
If you are not sure which LAN card is installed in your GC:
1.
2.
3.
To set the IP address locally
Caution
Be sure the IP address you enter is correct and is not in use by another device
(such as a printer).
1.
2.
Released: March 2004
Control modes
BootP
b
BootP
DHCP, GC front panel, Control Module
Turn the GC off.
Press and hold LOAD and turn the GC on. Continue to hold LOAD until
five dots appear in the display.
When the GC finishes initializing, you will see one of the following:
DHCP MODE:
DISABLED
Lantronics card installed
If the GC does not display either of these two screens, your GC uses card
J2552B or card J4100A and firmware that does not support local IP
addressing. If this is the case, BootP must be used to obtain the IP address.
Turn the GC off.
Press and hold LOAD and turn the GC on. Continue to hold LOAD until
five dots appear in the display.
6850 Series II GC User Information
a
a
, GC front panel
or
Setting your IP address
Table 38
lists the types of
GC firmware
A.03.XX or greater
A.03.XX or greater
A.03.XX or greater
IP ADDRESS
XXX.XX.XX.XX
Card J4100A installed
Installation
page 245 of
256

Hide quick links: