Procedure
Warning
To avoid a potential shock hazard when using liquid detection fluid, turn the
GC off and disconnect the main power cord. Be careful not to spill leak solu-
tion on electrical leads, especially the detector and inlet heater leads.
1.
2.
3.
Leak testing the cool on-column inlet
There are numerous places in the inlet that can leak. This procedure lets you
determine, in general, if there is an unacceptable leak in the inlet. If the inlet is
leaking, use an electronic leak detector to pinpoint the leaking component.
If you have access to a Control Module, you can perform this test
automatically. See its documentation for details. The Control Module test pro-
gram will prompt you to complete the hardware tasks outlined below.
Materials needed:
Procedure
Warning
Be careful! The oven and/or inlet may be hot enough to cause burns.
This test requires either a Control Module or a ChemStation/Cerity Chemical to
edit setpoints.
Released: March 2004
Using the leak detector, check each connection you have made for leaks.
Correct leaks by tightening the connections. Retest the connections;
continue tightening until all connections are leak-free.
Cap the septum purge vent with a 1/8-inch Swagelok cap.
1/4-inch wrench
Gloves (if the inlet is hot)
1/8-inch brass nut, 5180-4103
1/8-inch Vespel/graphite blank ferrule, 0100-1372
Column nut, 5181-8830
Vespel/graphite blank ferrule, 5020-8294
6850 Series II GC User Information
Routine Maintenance: Inlets
Maintaining a cool on-column inlet
page 181 of