FID checkout

Scope
Use the following procedure to verify proper FID operation.
Parts and equipment required
All inlet types
6850 evaluation column, 30 m
FID/TCD performance evaluation (checkout) sample, part no. 18710–60170
Chromatographic-grade gases: helium, nitrogen, hydrogen, and air
Split/splitless and purged packed inlets
10
O-ring, part no. 5180–4182 (split/splitless) or 5080–8898 (purged packed)
Septum, part no. 5181–1263
Cool on-column inlet
With septum nut:
With duckbill septum
PTV inlet
10
Inlet adapter, GRAPHPACK-2M, part no. 5182-9761
Silver seal, for GRAPHPACK-2M, part no. 5182-9763
Glass liner, multibaffle, part no. 5183-2037
Teflon ferrule, part no. 5182-9748 (septumless head)
Microseal replacement, part no. 5182-3444 (if installed)
GRAPHPACK 3D ferrule, part no. 5182-9749
Released: March 2004
µ
L syringe, part no. 9301–0810 or equivalent
Needle, part no. 5182-0831, for manual syringe
Septum nut, part no. 19245-80521
Needle, fused silica, part no. 19091-63000
Syringe, for fused silica needle, part no. 9301-0658
Septum, part no. 19245-40050
µ
L syringe, part no. 9301–0810 or equivalent
6850 Series II GC User Information
×
×
0.320 mm
0.25 mm HP-1, part no. 19091Z–413E
Verifying Performance