Maintaining A Flame Ionization Detector (Fid) - Agilent Technologies 6850 Series II User Information

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

9.
10. Check that the signal offset value is < 30 after the instrument equilibrates.

Maintaining a flame ionization detector (FID)

Warning
Flame ionization detectors use hydrogen gas as fuel. If hydrogen flow is on and
no column is connected to the detector inlet fitting, hydrogen gas can flow into
the oven and create an explosion hazard. Detector fittings must have either a
column or a cap connected at all times.
FID gas flow paths
Figure 32
maximum gas flows.
Released: March 2004
Load a method that uses the TCD.
Prep Run
Ready for autoinject
Signal: TCD
Stop
Not Ready
shows the pneumatics design for the FID and
6850 Series II GC User Information
Maintaining a flame ionization detector (FID)
5.5
Load
Start
Run
Routine Maintenance: Detectors
Table 36
shows the