Maintaining a Microcell Electron Capture Detector
(µECD)
Figure 33
Warning tag
Figure 33. The µECD
Replacing the µECD capillary column adapter liner
To replace the µECD capillary column adapter liner:
1.
Cap
Liner
Capillary column
adapter
Figure 34. The µECD adapter
2.
Released: March 2004
is an overview of the µECD and its electrometer.
Remove the adapter cap.
Check the liner. Replace it if it is broken and reinstall the cap. The
indentation must be at the cap end of the adapter (See
6850 Series II GC User Information
Maintaining a Microcell Electron Capture Detector (µECD)
Indentation
Routine Maintenance: Detectors
Vent line
Perforated cover
Electrometer
Figure
34).