FAULT 265
FAULT 267
FAULT 269
Possible cause:
FAULT 270
FAULT 272
FAULT 274
Possible cause:
FAULT 275
FAULT 277
FAULT 279
Possible cause:
FAULT 280
FAULT 282
FAULT 284
Possible cause:
FAULT 289
Possible causes:
FAULT 290
Possible causes:
Released: March 2004
Inlet invalid PID
Detector invalid PID
Pneu aux invalid PID
Defective flow module.
The GC main board has determined that the inlet/detector/aux flow module
calibration setpoints are no longer valid. The module is unusable.
Inlet bad checksum
Detector bad checksum
Pneu aux bad checksum
Defective flow module
The inlet/detector/aux flow module's calibration setpoints are out of range and
are not valid.
Inlet bad fact cal
Det bad factory cal
Pneu aux factory cal
Defective flow module
The inlet/detector/aux flow module's calibration is invalid.
Inlet I/O failure
Det I/O failure
Pneu aux I/O failure
Defective flow module.
The inlet/detector/aux flow module is malfunctioning. The GC main board
could not read/write to the module EEPROM.
Oven fan motor
Oven fan motor not connected/missing
Oven fan motor defective
Defective main board
The controller for the primary oven fan motor detected a fault, and the oven is
shut down for safety.
Zones not updating
Corrupted firmware
Defective main board
The program that controls one or more of the thermal zones has stopped, and
the zones may be uncontrolled. To prevent a thermal runaway, all thermal
zones are turned off.
Turn the GC off, then on. If the fault remains, update the GC firmware or
replace the main board.
6850 Series II GC User Information
Messages
Fault messages