Agilent Technologies 6850 Series II User Information page 63

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

2.
3.
4.
Table 18. FID Checkout Conditions (Method FIDCKO)
Column and sample
Type
Sample
Injection volume
Inlets
Temperature
Inlet pressure
Split/splitless
Mode
Purge flow
Purge time
PTV
Mode
Inlet temperature
Initial time
Rate 1
Final temp 1
Final time 1
Rate 2
Final temp 2
Final time 2
Purge time
Purge flow
Detector
Temperature
H
2
Air flow
Makeup flow (N
Lit offset
Released: March 2004
Enter the parameter values listed in
Define the column.
Save the method. We suggest using the name FIDCKO.
flow
)
2
6850 Series II GC User Information
Table
HP-1, 30 m × 0.32 mm × 0.25 µm part number 19091Z–413E
FID/TCD checkout 18710-60170
1 µL
250°C Purged packed and split/splitless
Oven Track (cool on-column)
40°C PTV (see below)
25 psi (constant pressure, helium)
Splitless
60 mL/min
0.75 min
Splitless
40°C
0.1 min
720°C/min
350°C
2 min
100°C/min
250°C
0 min
0.75 min
60 mL/min
300°C
30 mL/min
400 mL/min
25 mL/min
Should be < 20 pA
Verifying Performance
18.
FID checkout
page 63 of
256