Agilent Technologies 6850 Series II User Information page 47

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

6.
7.
Installing an adapter in a detector fitting
This is a general procedure for installing many types of adapters onto detector
fittings. See
1.
2.
3.
Released: March 2004
Insert the spacer into the vise-held Swagelok fitting.
The male Swagelok fitting and spacer should be kept on hand to be used
whenever new ferrules are being installed on a column.
Table 15
for adapter part numbers.
Assemble a brass nut and a graphitized Vespel ferrule onto the adapter.
Graphitized Vespel ferrule
Nut
Adapter
Insert the adapter straight into the detector base as far as possible. Hold
the adapter in this position and tighten the nut finger tight.
1/4-in. column: tighten an additional 3/4 turn with a 9/16-in. wrench
1/8-in. column: tighten an additional 1/4 turn with a 7/16-in. wrench
Detector fitting
Install your column. See
6850 Series II GC User Information
Adapter
"Installing packed metal columns" on page
Columns and Traps
Installing packed metal columns
48.
page 47 of
256