The 6850 Automatic Injector - Agilent Technologies 6850 Series II User Information

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

The 6850 automatic injector

Loading the turret
You can place one vial in the standard injector turret, or you can place up to
three vials in it by installing optional inserts (part no. 07673-40150). If you use
the eight sample position turret, you can place up to eight sample vials in it. See
Figure
Solvent A
Waste A
Solvent B
Waste B (or Sample 3*)
Standard turret
*Requires optional insert (part no. 07673-40150) to hold sample vial.
Figure 48.
Vial placement in the turret
Changing the turret
Your injector comes with two types of turrets. You may use either turret.
If you need to remove or replace your turret, use the following instructions for
proper replacement.
1.
2.
Released: March 2004
48.
(or Sample 2*)
Disconnect the injector cable from the GC.
Open the injector door. Unscrew and remove the knurled nut from the top
of the turret. See
Figure
6850 Series II GC User Information
Solvent A
Sample 1
Solvent B
Waste
Solvent B2
Sample 8
49.
Automatic Liquid Sampler and Injector
The 6850 automatic injector
8 position turret
page 236 of
Sample 1
256

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents