Agilent Technologies 6850 Series II User Information page 212

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

6.
Gas fittings
Remove front screws
7.
8.
9.
10. Loosen the back screw, remove the two front screws, and pull out the flow
To install the new FPD flow module
1.
2.
3.
4.
5.
6.
7.
8.
Released: March 2004
Remove the gas fittings from the flow module.
Unscrew the manifold block fitting from the flow module.
Loosen the two screws securing each of the connector covers, and slide
the connector covers off.
Push the two tabs to release the ribbon cable connector and disconnect it.
module from the lid.
Place the module in the lid.
Connect the ribbon cable connector.
Fasten the flow module to the lid using the three T–20 Torx screws.
Slide the connector covers into position and tighten the mounting screws.
Install the manifold block fitting and reconnect the gas fittings.
Reconnect the supply gases and turn them on. Reconnect the power cord.
Turn the GC on.
Load a method that uses the FPD.
6850 Series II GC User Information
Routine Maintenance: Detectors
Maintaining the flame photometric detector (FPD)
Slide covers off
page 212 of
256

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents