Agilent Technologies 6850 Series II User Information page 66

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Chemical checkout
1.
2.
3.
4.
pA
2000
1800
1600
1400
1200
1000
800
600
400
200
0
Figure 9. Sample FID checkout chromatogram
Released: March 2004
Press Prep Run to prepare the inlet for splitless injection. When "Ready
for manual inj" appears, inject 1 µL of the checkout sample and press
Start on the GC.
The chromatogram should be similar to
The acceptance criteria are:
Area counts for components labeled C14, C15, and C16 are all > 4,000.
The area counts ratio calculated as C14/C16 is 1.00 ± 0.05.
MDL (C16, pg/mL) = (436,000 × peak-to-peak noise)/(Area C16) ≤ 4.
If these criteria are not met, repeat the test. If after repeated testing the
criteria cannot be met, consult your service and user documentation for
additional information.
2
6850 Series II GC User Information
Figure
4
6
Time (min.)
Verifying Performance
FID checkout
9.
8
page 66 of
256