Agilent Technologies 6850 Series II User Information page 31

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Installing capillary columns in the TCD detector
1.
Detector fitting
Front ferrule
Back ferrule
Column nut
Column
2.
3.
4.
Table 13. Ferrules for the TCD Detector
Column outside diameter
0.8 mm
0.53 mm
0.45 mm
No-hole ferrule
Installing capillary columns in the FPD
The FPD uses an adaptable fitting that can use both packed and capillary col-
umns. If your adaptable fitting does not have a capillary adapter installed,
begin with step 1. If the capillary adapter is already installed in your adaptable
fitting, begin with step 5.
Released: March 2004
Assemble the ferrules and a 1/8-inch brass nut on the column as shown.
See
Table 13
for the proper ferrules. Trim off a short piece of column to
remove any ferrule fragments inside the column.
Insert the column into the detector until it bottoms. Do not attempt to
force it.
Slide the column nut and ferrule up the column to the detector and tighten
the nut finger tight.
Pull the column out 1 mm. Use a wrench to tighten the nut an additional
1/4 turn. The column should not move.
6850 Series II GC User Information
Back ferrule
G1530–80400
G1530–80400
G1530–80400
G1530–80400
Columns and Traps
Installing capillary columns
Front ferrule
G1530–80410
G1530–80420
G1530–80430
G1530–80440
page 31 of