7.
Gas fitting
Remove front screws
8.
9.
10. Push the two tabs to release the ribbon cable connector and disconnect it.
11. Loosen the back screw, remove the two front screws, and pull out the flow
12. Disconnect the TCD switching valve connector at the jumper.
To install the new TCD flow module
1.
2.
3.
4.
5.
6.
7.
8.
Released: March 2004
Remove the gas fitting from the flow module.
Unscrew the manifold block fitting from the flow module.
Loosen the two screws securing each of the connector covers, and slide
the connector covers off.
module from the lid.
Connect the TCD switching valve connector.
Place the module in the lid.
Connect the ribbon cable connector.
Fasten the flow module to the lid using the three T–20 Torx screws.
Slide the connector covers into position and tighten the mounting screws.
Install the manifold block fitting and reconnect the gas fittings.
Reconnect the supply gases and turn them on. Reconnect the power cord.
Turn the GC on.
6850 Series II GC User Information
Routine Maintenance: Detectors