Installing packed metal columns

There are two sizes of packed metal columns, 1/4-in. and 1/8-in., in common
use. This general procedure applies to both sizes of columns.
1.
2.
3.
Released: March 2004
Position the ferrule so that it is 17 mm from the column end. Place a small
mark (typewriter correction fluid is useful) at the back of the ferrule and,
making sure that the column is correctly positioned, insert the column end
into the adapter.
17 mm
Mark column here
Screw the column nut on finger tight. Using a 5-mm wrench, tighten the
column nut 1/8- to 1/4-turn. Be careful not to overtighten.
Check the connections for leaks. If there are any leaks at the column
adapter, tighten it slightly more with the open end wrench provided.
Install ferrules on your packed column (See
column" on page
44).
If the capillary column hanger is installed, remove it. (See
capillary column hanger" on page
See
Table 9
for fittings required. Install fittings as needed.
6850 Series II GC User Information
Installing packed metal columns
"Installing ferrules on a metal
22.)
Columns and Traps
"Removing the
page 43 of
256