Table 44.
Gas Purity Recommendations
Purity when used as carrier or
Gas
capillary makeup gas
Helium
99.9995%
Nitrogen
99.9995%
Hydrogen
99.9995%
Argon/Methane
99.9995%
Air (dry)
(Not used)
>44 cm (17.3 in)
Figure 56. 6850 GC bench space requirements
end
Released: March 2004
Purity when used as detector
support gas
99.9995%
99.9995%
99.9995%
99.9995%
Zero-grade or better
F8
F7
F6
m
i
F1
F2
F3
F4
F5
7
8
9
Esc
4
5
6
Enter
1
2
3
0
.
-
6850 Series II GC User Information
Summary of GC benchtop and lab requirements
>29.5 cm (12 in)
H:
50 cm (19.7 in)
61 cm (23 inch), FPD
W: 29 cm (11.4 in), standard
34 cm (13.4 in), cryo cooling
37 cm (14.6 in), 6850 ALS
D:
54 cm (21.3 in)
64 cm (25.2-inch), auxiliary
EPC
Wt: <23 kg (51 lb)
Site Prep Summary