Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

2.
3.
Caution
From this point on, handle the parts only with forceps or tweezers!
4.
5.
Reassembling the detector
Caution
Handle the clean collector and insulators only with forceps (or tweezers)!
Tools
T-20 Torx screwdriver
Warning
Shock hazard. Turn off the electrometer or the GC main power.
1.
2.
Lower collector assembly
3.
Released: March 2004
Use the nylon brushes to clean each piece.
Replace the pieces in the bath and clean again for 5 minutes.
Remove the pieces from the bath and rinse them thoroughly with hot tap
water and then with a small amount of methanol.
Place the pieces on a paper towel and let them air dry.
Confirm that the electrometer is turned off. If you are not sure, turn the GC
main power switch off.
Insert the lower insulator into the lower collector assembly. Install the
lower collector assembly and tighten the three screws.
Replace the collector and install the upper Teflon insulator.
6850 Series II GC User Information
Routine Maintenance: Detectors
Maintaining a flame ionization detector (FID)
page 197 of
256