Agilent Technologies 6850 Series II User Information page 196

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Top assembly
5.
Insulator
Collector
6.
Collector bottom
assembly
Cleaning the collector
1.
Released: March 2004
Lift out the collector. The upper insulator may be attached to the collector.
You may need to use the tweezers to grasp the collector.
Remove the three screws that hold the collector bottom assembly in place.
Lift off the assembly. Remove the lower insulator from the bottom
assembly. You may need to use the forceps to grab it.
Fill the ultrasonic cleaning bath with aqueous detergent, and place the two
insulators and the collector in the bath. Ultrasonically clean for 5 minutes.
6850 Series II GC User Information
Knurled nut
Routine Maintenance: Detectors
Maintaining a flame ionization detector (FID)
page 196 of
256