Agilent Technologies 6850 Series II User Information page 72

Gas chromatograph
Hide thumbs Also See for 6850 Series II:
Table of Contents

Advertisement

Figure 10. Sample µECD noise, wander, and drift plot
Chemical checkout
1.
2.
3.
4.
Released: March 2004
2.5
5
Press Prep Run to prepare the inlet for splitless injection. When "Ready
for manual inj" appears, inject 1 µL of the checkout sample and press
Start on the GC.
The chromatogram should be similar to
The acceptance criteria is:
Peak response ≥ 7500 × Noise
If these criteria are not met, repeat the test. If after repeated testing the
criteria cannot be met, consult your service and user documentation for
additional information.
6850 Series II GC User Information
detector
7.5
10
Figure
Verifying Performance
µECD Checkout
12.5
min
11.
page 72 of
256