Graphite and graphitized-Vespel ferrules
Place these ferrules in a petri dish and bake in the GC oven at 250 to 300°C for
30 minutes before use to remove organic compounds absorbed by the graphite.
Leave a petri dish of assorted ferrules in the GC oven to ensure a clean supply.
Vespel ferrules
These ferrules can be more leak-tight than graphite, but have a lower
temperature limit. They should be retightened after a few oven temperature
cycles to ensure a good seal. Be sure to use the correct ferrule for the size
column you are using.

Ferrules and O-rings for glass packed columns

Table 10
See the Agilent catalog for consumables and supplies for a more complete listing.
Ferrules that are prepared improperly cause leaks and contamination. To avoid
problems, place graphitized Vespel ferrules in a petri dish and bake in the GC
oven at 250 to 300°C for 30 minutes before use to remove organic compounds
absorbed by the graphite. Leave a petri dish of assorted ferrules in the GC oven
to ensure a clean supply.
Table 10. Glass Packed Columns Consumables
Item*
1/4-in. graphitized Vespel ferrule, pkg of 10
Silicone O-ring, 6.0-mm
* O-ring and ferrules ids
Released: March 2004
lists some of the nuts and ferrules used with packed glass columns.
6850 Series II GC User Information
Ferrules and O-rings for glass packed columns
Typical use
Inlet/Detector liners,
1/4-in. glass packed columns
1/4-in. glass packed columns
Columns and Traps
Part no.
5080-8774
0905-0322