Figure 17. Rotary sampling valve actuator air supply
Released: March 2004
Turn off the oven and valve box heater and allow the valve to cool to room
temperature.
Alternately, load the SERVICE method and wait until the GC becomes
READY.
Using a control module or data system, set the valve to the OFF position.
Turn off the air supply to the valve actuator, then loosen the connection to
the air supply tubing to relieve the back pressure at the actuator. When the
pressure is released, disconnect the supply tube from the valve box. See
Figure
17.
6850 Series II GC User Information
Routine Maintenance: General
Maintaining a sampling valve (rotary type)