Routine Maintenance: Detectors

The procedures described in this section can be performed by the user to
maintain the detectors in the GC. They may involve exposure to heated sur-
faces, but not to hazardous voltages.

Maintaining a thermal conductivity detector (TCD)

If the TCD is displaying problems such as a wandering baseline, increased
noise level, or changes in response on a checkout chromatogram, it is probably
contaminated with deposits from such things as column bleed or dirty samples.
The TCD is cleaned by a process known as thermal cleaning or bakeout.
Perform bakeout only after you have confirmed that the carrier gas and the
flow system components are leak and contaminant free.