10. Drift (for a 5-minute time window) should be ≤ |5| display units/hour. See
Figure 12. Sample FPD noise, wander, and drift plot
Chemical checkout
1.
2.
3.
4.
Released: March 2004
Figure
12.
Press Prep Run to prepare the inlet for splitless injection. When "Ready
for manual inj" appears, inject 1 µL of the checkout sample and press
Start on the GC.
The chromatogram should be similar to
The acceptance criteria are:
Sulfur peak response ≥ 24000 counts
Phosphorus peak response ≥ 24000 counts
If these criteria are not met, repeat the test. If after repeated testing the
criteria cannot be met, consult your service and user documentation for
additional information.
6850 Series II GC User Information
Verifying Performance
Figure
13.
FPD Checkout
page 79 of
256