Ieee-Common Commands - Keysight N777 C Series Programming Manual

Tunable laser source
Table of Contents

Advertisement

3
Instrument Setup and Status

IEEE-Common Commands

Command:
*CLS
Syntax:
*CLS
Description:
The Clear Status (*CLS) command clears the status byte by emptying the error queue and clearing all the event registers
(SESR) including the Data Questionable Event Register, the Standard Event Status Register, the Standard Operation Status
Register and any other registers that are summarized in the status byte.
Parameters:
none
Response:
none
Example:
*CLS
Command:
*ESE
Syntax:
*ESE<wsp><value>
0
Description:
The standard Event Status Enable command (*ESE) sets bits in the Standard Event Status Enable Mask (SESEM) that enable
the corresponding bits in the standard event status register (SESR).
The register is cleared:
at power-on,
by sending a value of zero.
The register is not changed by the *RST and *CLS commands.
Parameters:
The bit value for the register (a 16-bit signed integer value):
Bit
7 (MSB)
6
5
4
3
2
28
Common Commands
IEEE-common commands which can be used with the instruments. This
section gives fuller descriptions of each of these commands.
value
255
Mnemonic
Power On
Not Used
Command Error
Execution Error
Device Dependent Error
Query Error
on page 16 gave a brief introduction to the
Keysight N777xC Series Tunable Laser Family Programming Guide
Decimal Value
128
64
32
16
8
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7776cN7778cN7779c

Table of Contents