Low-Side Common Resistor Calibration -- R; Com,L - Keysight PD1550A System Installation Manual

Advanced dynamic power device analyzer
Table of Contents

Advertisement

3 Low-Side Common Resistor Calibration -- R

3 Low-Side Common Resistor Calibration -- R

Hardware Setup:
- Remove the DUT module from the Interface Module before calibration.
- Remove the Low-Side Gate Drive Module from the Interface Module.
- Install the Low-Side Gate Resistor Module that you want calibrated.
- Install a UDI cable on the Interface Module.
- Use two sets of calibration cables for a four-wire resistance measurement as
Figure 11 Low-Side R
Procedure:
Keysight PD1550A Installation Guide

COM,L

follows:
– Red cable from SMU Force to Low-Side COM terminal.
– Red cable from SMU Sense to Low-Side COM terminal.
– Black cable from SMU Force to Low-Side VG terminal.
– Black cable from SMU Sense to Low-Side VG terminal.
To Low-Side
COM terminals
Calibration Connections
COM
Click Start to perform the calibration and continue to next step.
PD1550A System Calibration- AutoCal
COM,L
To Low-Side
VG terminals
109

Advertisement

Table of Contents
loading

Table of Contents