Ac V (V) Averaging Sense Performance Verification Tests - Keysight U1242C Service Manual

Handheld digital multimeter
Hide thumbs Also See for U1242C:
Table of Contents

Advertisement

1
Calibration Procedures
Table 1-5

AC V (V) averaging sense performance verification tests

Test function
AC mV (mV) averaging sense
Turn the rotary switch to the
AC V (V) averaging sense
Turn the rotary switch to the
View
.
Shift
Esc
[a] The impedance may be set to >1 GΩ from the Setup mode; the default value is 10 MΩ in parallel with 100 pF (nominal).
[b] Before turning the rotary switch, you must enable the mV measurement function from the Setup menu first. For enabling the mV
measurement function from the Setup menu, refer to the U1241C/U1242C User's Guide.
28
position.
1
2
position and press
V
5520A/5522A
Range
output
100 mV, 40 Hz
[a][b]
100 mV, 1 kHz
100 mV
100 mV, 2 kHz
600 mV, 40 Hz
[a][b]
600 mV, 1 kHz
600 mV
600 mV, 2 kHz
1000 mV, 40 Hz
1000 mV
1000 mV, 1 kHz
1000 mV, 2 kHz
10 V, 40 Hz
10 V
10 V, 1 kHz
10 V, 2 kHz
100 V, 45 Hz
100 V
100 V, 1 kHz
100 V, 2 kHz
1000 V, 45 Hz
1000 V
1000 V, 1 kHz
1000 V, 2 kHz
Keysight U1241C/U1242C Service Guide
Error from nominal 1 year
U1241C
U1242C
±1.05 mV
N/A
±1.05 mV
±1.55 mV
±6.5 mV
N/A
±6.5 mV
±9.5 mV
±10.5 mV
N/A
±10.5 mV
±15.5 mV
±0.105 V
N/A
±0.105 V
±0.155 V
±1.05 V
N/A
±1.05 V
±1.55 V
±10.5 V
N/A
±10.5 V
±15.5 V

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1241c

Table of Contents