Operation Using The Batcher; Run A Batch - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20039472

12 Operation using the batcher

12.1 Run a batch

You can run a batch using any of the presets that are enabled. You can change the target for the current
batch. You can pause and resume a batch in process. You can allow the batch to complete normally or you
can end it manually.
Restriction
You cannot use the Field Communicator to run a batch.
Prerequisites
The preset that you want to use must be enabled.
Procedure
1. Select the preset you want to use.
Using the display: Choose Menu → Batcher → Run Batch and set Active Preset to the preset you
want to use.
Using ProLink III: Choose Device Tools → Configuration → Batcherand set Current Preset Number
to the preset you want to use.
The preset contains parameters that control batch processing.
2. If desired, change the configured target to the target to be used for this batch only.
Using the display: Choose Menu → Batcher → Run Batch and set Batch Target to the desired value.
Using ProLink III: Choose Device Tools → Application Control → Run Batcher and set Current
Target to the desired value.
You cannot set the target to a value that is greater than the Maximum Target value configured for the
batching application.
Changing the target here does not change the target configured in the preset.
3. Begin the batch.
Using the display: Choose Menu → Batcher → Run Batch → Batch Control and select Begin.
Using ProLink III: Choose Device Tools → Application Control → Run Batch and click Begin Batch.
The transmitter automatically resets the batch total to 0 if you have selected Reset on Start.
Otherwise, you must manually reset the batch total before the batch will start.
Reset on Start is not applicable for custody transfer batcher applications.
4. Monitor the batch during processing.
Using the display: Choose Menu → Batcher → Run Batch → Batch Control and observe the value
displayed in Actual Total.
Using ProLink III: Choose Device Tools → Application Control → Run Batcher and observe the
values displayed in Current Total and Batch Time.
Configuration and Use Manual
Operation using the batcher
June 2019
165

Advertisement

Table of Contents
loading

Table of Contents