GRAPHTEC midi LOGGER GL240 User Manual page 70

Hide thumbs Also See for midi LOGGER GL240:
Table of Contents

Advertisement

CHAPTER 3 Settings and Measurement
(1)-7 Logic/Pulse
This is used to select the processing method for digital input.
Selection item
Off
Logic
Pulse
(1)-8 Input
This is used to set the pulse measurement mode. This setting is available only if Pulse is selected in (1)-7.
Selection item
Off
Revol.
Counts
Inst.
(1)-9 Filter
This is used to set the filter for input.
Selection item
Off
On
(1)-10 Pulse Slope
This is used to set the slope (direction) to count the number of pulses. This setting is available only if Pulse
is selected in (1)-7.
Selection item
H
L
Allocation of CH number of logic and pulse data is as shown below.
MSB
15 14 13 12 11 10 9
3-28
Digital input measurement is disabled.
Digital input is processed as logic signals.
Digital input is processed as pulse signals.
Pulse input measurement is disabled.
The number of pulses per sample interval is counted and converted to the number of revolutions per
minute.
Captures the cumulative number of pulses for each sampling interval from the start of measurement.
Captures the number of pulses for each sampling interval.
Disables hardware filter.
Enables hardware filter. It is effective in a noisy environment.
The filter is approximately 30 Hz (-3 dB).
Counts the rising edges of pulses.
Counts the falling edges of pulses.
LSB
8
7
6
5
4
3
2
1
Bit 0 to 3: Main unit data
Bit 4 to 7: Wireless sensor data
Description
Description
Description
Description
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents