Connecting The Signal Input Cables - GRAPHTEC midi LOGGER GL240 User Manual

Hide thumbs Also See for midi LOGGER GL240:
Table of Contents

Advertisement

CHAPTER 2 Checks and Preparation

2.5 Connecting the Signal Input Cables

This section describes how to connect the signal input cables.
During wiring, confirm that the signal's supply source is turned OFF to prevent electrical shocks.
Also, position the GL240 input cable away from any power lines and ground cables.
Terminal Configuration and Signal Types
Terminal assignment of standard terminal and Withstand high-voltage high-precision terminal is common.
CH1
+
-
Connection diagram
DC voltage input
Voltage input
Resistance temperature detector input
not available
+ .....................................High-voltage terminal (terminal for high-voltage input signals)
- ......................................Low-voltage terminal (terminal for low-voltage input signals)
Input configuration
Measurement range
Thermocouples
A/D resolution
Filter
2-6
CH10
+
-
Item
Isolated input, scanning
20, 50, 100, 200, 500 mV/F.S.; 1, 2, 5, 10, 20, 50, 100 V/F.S.; 1-5V
K, J, E, T, R, S, B, N, W (WRe 5-26)
16-bit (Effective resolution: Approx. 1/40,000 of the +/- range)
Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the set sampling count is used.
If the sample interval exceeds 5 seconds, the average value of data
obtained in a sub-sample (5 seconds) is used.
Thermocouple input
Compensationcopper wire
Current input
+
DC current
-
Shunt resister
Ex: The current is converted to the voltage in the shunt register.
For 4 to 20mA current input, installing 250 ohms (0.1%)
resister for converting 1 to 5V.
* 250 ohms shunt register is the option B551.