Thermal channel commands
E5080 Series Cryogen Monitor User Guide
Command
XIY
X1N
X2Y
X2N
CNAnn
CNBnn
T1xxy
T2xxy
CEAnnkkkRRRRR
CEBnnkkkRRRRR
Function
Switch Thermal channel A on
Switch Thermal channel A off
Switch Thermal channel B on
Switch Thermal channel B off
Set number of entries in thermal sensor data table
channel A to nn+1
Set number of entries in thermal sensor data table
channel B to nn+1
Set thermal channel A alarm level where xx.y is in
degs K,
Set thermal channel B alarm level where xx.y is in
degs K,
Set thermal sensor table A entry
where nn is table index value starting from 00
For software versions 1 to 4 where kkk is x.xx degs K
For software versions 5 onward where kkk is xx.x
degs K
where RRRRR is Ohms with leading zeroes
Set thermal sensor table B entry
where nn is table index value starting from 00
For software versions 1 to 4 where kkk is x.xx degs K
For software versions 5 onward where kkk is xx.x
degs K
where RRRRR is Ohms with leading zeroes