Thermal Channel Commands - Agilent Technologies E5080 Series User Manual

Cryogen monitor with pressure control
Table of Contents

Advertisement

Thermal channel commands

E5080 Series Cryogen Monitor User Guide
Command
XIY
X1N
X2Y
X2N
CNAnn
CNBnn
T1xxy
T2xxy
CEAnnkkkRRRRR
CEBnnkkkRRRRR
Function
Switch Thermal channel A on
Switch Thermal channel A off
Switch Thermal channel B on
Switch Thermal channel B off
Set number of entries in thermal sensor data table
channel A to nn+1
Set number of entries in thermal sensor data table
channel B to nn+1
Set thermal channel A alarm level where xx.y is in
degs K,
Set thermal channel B alarm level where xx.y is in
degs K,
Set thermal sensor table A entry
where nn is table index value starting from 00
For software versions 1 to 4 where kkk is x.xx degs K
For software versions 5 onward where kkk is xx.x
degs K
where RRRRR is Ohms with leading zeroes
Set thermal sensor table B entry
where nn is table index value starting from 00
For software versions 1 to 4 where kkk is x.xx degs K
For software versions 5 onward where kkk is xx.x
degs K
where RRRRR is Ohms with leading zeroes