If Errors Occur When You Perform Self-Calibration Or Diagnostics - Agilent Technologies 4156b User Manual

Precision semiconductor parameter analyzer
Hide thumbs Also See for 4156b:
Table of Contents

Advertisement

If You Have A Problem
If Errors Occur
If Errors Occur when You Perform Self-calibration or
Diagnostics
The following are the error codes that are displayed at the bottom of the SYSTEM:
SELF-CALIBRATION/DIAGNOSTICS screen if errors occur when you perform
self-calibration or diagnostics.
If errors occur, write down the displayed error codes and contact the nearest Agilent
Technologies Sales and Service office. Up to seven error codes can be displayed at
the bottom of the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS screen. To
display the error codes, move pointer to a desired test item.
Error codes for measurement unit
The following are the error codes for measurement units. Error codes are 5-digit
numbers.
1xxyy
xx: measurement unit
00: VSU1,2 and VMU1,2
01 to 06: SMU1 to SMU6
07: PGU1,2
08: GNDU
09: AD converter
yy: error number
AD converter failed ROM or RAM self-test. Measurement unit failed
1xx05
AD converter test as a pretest for calibration or self-test.
Successive approximation AD converter failed. Measurement unit
1xx06
failed AD converter test as a pretest for calibration or self-test.
Integrating type AD converter failed. Measurement unit failed AD
1xx07
converter test as a pretest for calibration or self-test.
1xx08
AD converter test reached timeout. Measurement unit failed AD
converter test as a pretest for calibration or self-test.
Overvoltage occurred for a measurement unit.
1xx11
1xx12
Overcurrent occurred for a measurement unit.
8-14