DESCRIPTION OF THE N-20PA PORTABLE PULSE OXIMETER

1.1 Manual Overview
1.2 Warnings, Cautions, and Notes
1.3 Description of the N-20PA Portable Pulse Oximeter
This manual contains information for servicing the N-20PA portable pulse
oximeter. Only qualified service personnel should service this product. Before
servicing the device, read the operator's manual carefully for a thorough
understanding of its operation.
This manual uses three terms that are important for proper operation of the
device: Warning, Caution, and Note.
A warning precedes an action that may result in injury or death to the patient or
user. Warnings are boxed and highlighted in boldface type.
A caution precedes an action that may result in damage to, or malfunction of, the
device. Cautions are highlighted in boldface type.
A note gives information that requires special attention.
The Nellcor
N-20PA portable pulse oximeter provides noninvasive
measurement of, and continuous information about, the percent of oxygenated
hemoglobin compared with total hemoglobin and pulse rate. A pulse amplitude
indicator provides a qualitative indication of pulse activity and patient perfusion.
Patients are connected to the instrument by a Nellcor sensor. The sensor LEDs
are driven by the SpO
2
signals, and provides CPU-adjustable gain stages. The N-20PA CPU measures
the sensor's analog outputs, continually controls the signal gain, and calculates
SpO
and pulse rate.
2
The N-20PA is automatically calibrated each time it is switched on, and
whenever a new sensor is connected; it sets sensor-specific calibration
coefficients by reading a calibration resistor in the sensor. Also, the intensity of
the sensor's light sources are adjusted automatically to compensate for
differences in tissue thickness and darkness.