Bunnell Life Pulse High Frequency Ventilator Circuit - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Circuit Connection
Diagrams
Connection to a Bunnell Life Pulse
High Frequency Ventilator Circuit
WARNING:
• The INOmax DS
be used with the Bunnell Life Pulse as ventilator flow rates
are normally below the recommended ventilator flows.
• Place the Life Pulse in Standby prior to suctioning the
patient to avoid NO delivery transiently exceeding the
set dose by up to 30 ppm. Press ENTER to reestablish
ventilation as soon as the catheter is removed from the
airway. This will limit the extent of over delivery above the
NO set dose.
Caution:
• If the set dose is below 5 ppm and the Servo pressure is 2.0 psig.
or less, this will result in flow rates outside of the specification of
the Injector Module and fluctuating NO values may result.
• A one-way valve should be placed between the injector module
and the humidifier chamber to prevent water from backing up into
the injector module if the Life Pulse is either put into Standby or
cycled OFF.
• There are higher pressures in the breathing circuit than normal;
use only parts provided in disposable package #50046 and tightly
secure all connections.
Part No. 20728 Rev-01
2012-09
backup mode (250 mL/min.) should not
IR
28