Inomax Ds Ir Plus Mri Cart Operation - Mallinckrodt INOmax DSIR Plus MRI Operation Manual

Table of Contents

Advertisement

INOmax DS
IR
• Always verify that the INOmax DS
WARNING:
positioning in the MR scanner room.
• Always verify that the INOmax DS
attached to the cart.
• Verify at least one gauss alarm is functioning properly prior to use. Do not use
the INOmax DS
• The gauss alarm will sound if the INOmax DS
the MR scanner. If alarm sounds, move system away from the MR scanner until
the gauss alarm stops sounding.
• If the cart fails to move when the brake handle is pulled or moves when the
brake handle is not pulled, do not use the INOmax DS
your local representative.
Note:
• The INOmax DS
cart when in the scanner room.
• To move the INOmax DS
handle.
• Two gauss alarms are attached to the cart and will sound if the cart is moved too close to the
scanner bore.
• The tether attachment point allows for a facility supplied cable to be secured to the cart as a
redundant means to limit the distance the cart can move.
1-4
Plus MRI Cart Operation
Plus MRI if neither gauss alarm is functional.
IR
Plus MRI cart provides an auto-brake to prevent inadvertent movement of the
IR
Plus MRI cart, pull up on the auto-brake handle, toward the cart
IR
Plus MRI auto-brake is engaged after
IR
Plus MRI and INOblender are securely
IR
Plus MRI system is too close to
IR
Plus MRI and contact
IR
Part No. 20568 Rev-02
2015-08