Calibration Error Codes - Thermo Scientific Orion 2117HL User Manual

High level chloride analyzer
Table of Contents

Advertisement

Calibration
Calibration
Error Codes
Error
Operator Action
Stability Time Out Error
Error
Press
to continue the calibration
41
despite the reading instability.
or
Press
to abort the calibration and
refer to Exit 3. Repeat the calibration or
refer to the troubleshooting section.
Calibration Errors
Error
Press
to abort the calibration and
42
refer to Exit 3.
Verify that the correct standards were
used during the calibration, the standard
values were entered correctly and in the
right order and the electrodes are working
properly and then repeat the calibration.
Error
Press
to abort the calibration and
107
refer to Exit 3.
Error
Press
to continue the calibration
109
despite the bad slope.
or
Press
to abort the calibration and
refer to Exit 3.
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
At any point during a calibration, the appropriate error message will be
displayed if there is a calibration related error.
The following table lists common calibration error codes with their
marquee display and required operator actions.
Scrolling Marquee
READINGS WERE UNSTABLE
PRESS ENTER TO CONTINUE
CALIBRATION OR PRESS EXIT
TO ABORT
INVALID SLOPE
PRESS ENTER
CALIBRATION DATA TOO
CLOSE TOGETHER
PRESS ENTER
CH1 BAD SLOPE
PRESS ENTER
Main Display
Notes
E041
The required system
stability was not
reached and a time
out error occurred.
Perform electrode
cleaning, conditioning
and troubleshooting.
E042
The calculated slope
was invalid.
Standard values were
likely entered in the
wrong sequence
or wrong standard
values were entered.
E107
Use new standards.
Perform electrode
cleaning, conditioning
and troubleshooting.
E109
The calculated slope
was outside of the
recommended range.
Perform electrode
cleaning and
conditioning, refer to
the troubleshooting
section and repeat the
calibration.
IV-11