4.2.3

Entering treatment parameters

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
A parameter field (green background) can be
selected by turning the rotary selector.
Turning the rotary selector to the right will
select the menu fields on the right down to the
Venous bubble catcher level field. This field
remains selected even if you turn the rotary
selector further.
Turning the rotary selector to the left will
select the menu fields on the left up to the
Substituate parameter. This field remains
selected even if you turn the rotary selector
further.
The procedure for entering treatment
parameters will be explained starting with the
example of the Blood flow.
After the selected Blood flow parameter has
been confirmed with [OK], the currently
selected actual value will be shown in black in
the first line of the parameter field and the
value to be changed (target value) will be
shown in white in the second line.
Turning to the left will reduce, turning to the
right will increase the value.
If the target value is turned to the full scale
value, it will be displayed in red. This applies
to both the lowest and the highest scale
value.
While changing the value, the actual value
will still be displayed in black. The new target
value is shown in white below the actual
value.
Chapter 4: Operation
4-7