Equipment Storage And Maintenance - Fresenius Medical Care 2008K Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Appendix B

Equipment Storage and Maintenance

Follow the storage and maintenance procedures for dialysis equipment used for Intermittent
Hemodialysis (IHD) in the ICU setting.
Warning! Possible explosion hazard if used in the presence of flammable anesthetics.
Storage Location
If the dialysis unit is incorporated in the hospital, the equipment should be stored so that it
will not be damaged. The maintenance of the equipment is normally part of the duty of the
dialysis service technician. Depending on the disinfectant and the storage time, frequent
flushing of the equipment is necessary.
If it is more convenient to store the equipment close to the ICU, the dedicated storage space
should have an access to tap water, power and a drain. The room should be well vented, if
disinfectant is used.
Storage Preparation
Before storing the 2008K hemodialysis machine, the hydraulics should be disinfected. It is
also necessary to wipe the external parts of the machine with a surface cleaner. Frequency of
disinfection and length of the dwell time depend on the disinfectant and shall be determined
by acceptable culture result. The following table lists commonly used procedures to store
equipment and then return it from storage. Validate your own procedure according to the
facility's policy. After prolonged storage, use a bleach disinfectant prior to patient use.
Table 34 – Disinfectant Information
Disinfectant
Formaldehyde
Bleach
Renalin
Heat
216
Program
Fill Hydraulic
(Service Mode)
Chemical Disinfection
program
Chemical Disinfection
program
Heat Disinfection
program
2008K Hemodialysis Machine Operator's Manual Rev P
Dwell time
Retest and Repeat
Unlimited
3 – 4 weeks
Only for time of
24 hours
disinfection program
Only for time of
24 hours
disinfection program
Recirculate and shut
24 hours
machine off

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents