Chapter 3—Setting Treatment Parameters

New Treatment Key

When the 2008K hemodialysis machine is first turned on in preparation for daily operation
(after a long power down), all treatment parameters revert to their default settings. This can
also be accomplished by pressing the New TX (New Treatment) key when in Dialysis Mode.
The New TX key allows the operator to reset patient treatment parameters to their default
settings without interrupting the power to the machine. This must be done in preparing the
2008K hemodialysis machine for all subsequent treatments after the first of the day.
After pressing the New TX key or after a long power down:
All treatment data (blood pressure, Kt/V) are deleted. The treatment summary
information is moved to the previous record in the "Trends" screen.
The RTD counter is reset to zero
All heparin treatment parameters are reset to zero
SVS profile is reset to None
UF treatment parameters are reset as follows:
UF profile is reset to None
UF Removed is reset to zero
UF Goal = 3000
UF Time = 3:00
UF Rate = 1000
The "Dialysate" screen is displayed and the concentrate will need to be confirmed
To activate the New Treatment Option:
1) Press
New TX
Figure 25 – New TX key
58
2) Press
CONFIRM
2008K Hemodialysis Machine Operator's Manual Rev P