T- 1.0.6 OPTIONAL BLOOD TEMPERATURE MODULE (BTM) EQUIPPED?
Per the Figure right, TWO (2) possible
scenarios:
1)
IF NOT equipped with the optional
Blood Temperature Module (BTM): See
procedure number
2)
IF equipped with a Blood Temperature
Module (BTM): See parts a THROUGH
g below:
a) Place the machine into Service Mode → Options → Hardware Options.
b) Next to 'BTM', place the 'X' into the "No" box.
c) Press 'Enter' to turn the "X" blue!
To prevent damage turn the machine OFF!
d)
d) To avoid pulling cables loose, GENTLY open the card cage.
e) The Functional Board is the 3
ribbon cable from the P8 connector.
Close the card cage!
f)
g) Where was the problem ORIGINALLY occurring, 1) In Dialysis Program i.e. connected to
concentrate? OR 2) In Heat Disinfect? TWO (2) possible scenarios:
1)
IF (and ONLY if) in Dialysis Program
239).
2) IF in Heat Disinfect
(2) possible scenarios next page:
T- 1.0.7
(page 239).
rd
circuit board from the right! Figure below, unplug the ten (10) pin
: Return to Heat Disinfect and allow up to thirty (30) minutes. TWO
2008T Troubleshooting Guide
P/N 490292 Rev. A
: See procedure number
T- 1.0.7
(page
Page 237