3
Components Installation
Installing Valves and Fittings into the Main Oven

Installing the Rotary Valves into the Main Oven

40
TRACE 1310 Auxiliary Oven Instruction Manual
To install a rotary valve into the main oven
Installing valves consists of the following steps:
• Installing the solenoid valves
• Disassembling the rotary valve and actuator
• Installing the rotary valve and standoff
• Connecting gas lines from the solenoid to the actuator
• Installing the actuator to the valve standoff
CAUTION Turn the actuator gas off before you start these procedures.
1. Installing the solenoid valves.
a. Install the solenoid manifold on the floor of the unheated compartment.
See the example of
Figure 52. Rotary Valve Installation (1)
b. Remove the rectangular-shaped cap from the manifold using the phillips screwdriver.
c. Secure the solenoid valve to the under plate using the two screws supplied with the
valve. See
Figure
CAUTION Tighten the screws evenly, taking care that the solenoid gasket rests on all
surfaces to prevent leaks.
Figure
52.
53.
Solenoid Manifold
Thermo Scientific