Flow - Thermo Scientific TRACE 1300 Instruction Manual

Auxiliary ovens
Hide thumbs Also See for TRACE 1300:
Table of Contents

Advertisement

3
Components Installation
Installing the Auxiliary Gas Modules
76
TRACE 1310 Auxiliary Oven Instruction Manual
Figure 110. Pneumatic Circuit
IN
AUXILIARY GAS
VENT
To install the split fitting option
This operating sequence provides instruction for installing the split option kit P/N 19071018
into the TRACE 1310 Auxiliary Oven.
1. Power off the GC and the TRACE 1310 Auxiliary Oven.
2. Remove the L-shape cover of the Auxiliary Oven.
3. Remove the L-shape cover of the main oven to access its interior.
4. Modify the gas inlet port of the channel for controlling split flow by plugging the gas inlet
fitting with the provided 1/8-in.cap.
5. Replacing the M5 cap with a M5 gas inlet fittings. See the example in
AUXILIARY OVEN (Main Oven)
1
2
6
6
Sampling
Sample
Loop
5
5
3
4
4

FLOW

RESTRICTOR
1 m; 0.5 mm i.d.
CHARCOAL TRAP
To DETECTOR
COLUMN
SPLITTER
SPLIT
Figure
111.
Thermo Scientific

Hide quick links: