Thermo Scientific TRACE 1300 Instruction Manual page 88

Auxiliary ovens
Hide thumbs Also See for TRACE 1300:
Table of Contents

Advertisement

3
Components Installation
Installing Valves and Fittings in the Unheated Compartment
62
TRACE 1310 Auxiliary Oven Instruction Manual
a. Install the solenoid manifold on the rear wall of the valve compartment.
See the example of
Figure 88. Unheated Diaphragm Valve Installation (3)
b. Remove the rectangular-shaped cap from the manifold using the #0 phillips
screwdriver.
c. Secure the solenoid valve to the under plate using the two screws supplied with the
valve. See
Figure
89.
CAUTION Tighten screws evenly, taking care that the solenoid gasket rests on all surfaces
to prevent leaks.
Figure 89. Unheated Diaphragm Valve Installation (4)
d. Connect the cable from the solenoid valve to the proper timed event output jack of
the Events connector marked J24 on the VOBP-HRM board located over the power
section. See
Figure
Figure
88.
90.
Solenoid Manifold
Thermo Scientific