Thermo Scientific TRACE 1300 Instruction Manual page 156

Auxiliary ovens
Hide thumbs Also See for TRACE 1300:
Table of Contents

Advertisement

5
Maintenance & Troubleshooting
Troubleshooting
130
TRACE 1310 Auxiliary Oven Instruction Manual
Table 9. Methanizer Analytical Troubleshooting
Symptom
No peaks at all.
CO or CO
peaks not present in the
2
chromatogram.
Diagnosis
Column broken or disconnected.
Defective electrometer or
amplifier.
FID flame is out.
Poor or missing electrical
connections.
Methanizer temperature not
correct.
Catalyst exhausted.
Remedy
Check the column and
connections.
Replace the electrometer or the
amplifier.
Light the flame.
Check the cables connections.
Set the correct temperature.
Replace the metal tube containing
the catalysts.