4. Installing the solenoid valves.
a. Install the solenoid manifold for the actuation of the valves on the floor of the
unheated compartment. See the example of
Figure 96. Unheated Valve for Liquid Samples Valve Installation (5)
Note The solenoid manifold for the liquid sample valve could have two solenoid
valves already mounted. See
interest.
b. Remove the rectangular-shaped cap from the manifold using the phillips screwdriver.
c. Secure the solenoid valve to the under plate using the two screws supplied with the
valve See
Figure
d.
CAUTION Tighten screws evenly, taking care that the solenoid gasket rests on all surfaces
to prevent leaks.
Figure 97. Unheated Valve for Liquid Samples Valve Installation (6)
d. Connect the cables from the solenoid valves to the proper timed event output jack of
the Events connector marked J24 on the VOBP-HRM board located over the power
section. See
Figure
Installing Valves and Fittings in the Unheated Compartment
Solenoid Manifolds
Figure 17
on
page
98.
TRACE 1310 Auxiliary Oven Instruction Manual